Product Introduction Apply in wards, ICU rooms, rescue rooms and operation recovery
rooms etc. Technical Characteristics Maximum oxygen supply: 60m3/h Terminal output pressure: 0.1-0.6MPa (adjustable) Oxygen-supply terminal is installed on terminal curing strip with
absorption terminal. Terminal curing strip is installed on the wall
of bed head, 1.3~1.5m from floor. Strong electricity, weak electricity and separation set in air
pipeline adopted in internal terminal curing enhance the security
of the whole unit. Headlight, power socket and talk extension adopt built-in
installing mode, the whole figure beautiful and elegant. Lock type oxygen repair switch in terminal unit (only special key
for the lock) not only can ensure continuity of terminal oxygen
supply, but also cut off air source when plug and draw valve need
to repair.